REFRACTORIES & LIME ›› 2021, Vol. 46 ›› Issue (3): 22-25.

Previous Articles     Next Articles

Comparison of the applicable standard for quantitative analysis of α-quartz by X-ray diffraction before and after revision

Wang Fuyun   

  1. Sinosteel Luonai Materials Technology Corporation, Luoyang 471039, China
  • Online:2021-06-10 Published:2021-07-12

Abstract:

The standard for quantitative analysis of α-quartz phase by X-ray diffraction has always been YB/T 172-2000.The revised standard will be adopted on April 1,2021.The comparison of the standard before and after revision shows that when the content is ≤5.2%,the absolute error before and after revision is ≤0.10;When the content is >5.2%,the absolute error before and after revision is >0.10;and the absolute error increases with the increase of the content.The revised standard has well inherited and developed the technical requirements of YB/T 172-2000.It has strong operability,high accuracy,good repeatability,and it is more convenient and fast.The diffraction intensity value of the sample can be directly substituted into the working curve to obtain the analysis results.

Key words: X-ray diffraction, Quantitative analysis, α-quartz, Method comparison